Infection Control Standards

Deficiencies: All deficiencies related to Infection Control must be addressed in a corrective action plan. 

1) Written site-specific policy/procedures or manufacturer’s instructions for instrument/equipment sterilization are available to staff. 

If site uses either cold chemical sterilization or an autoclave, site shall have specific policy/procedures or manufacturer’s instructions addressing instrument pre-treatment, preparation, machine loading, operation safety precautions,and other required processes that are available to staff to follow. 

Staff adheres to site-specific policy and/or manufacturer/product label directions for the following procedures: 

VI.D.2) Cleaning reusable instruments/equipment prior to sterilization. 

Cleaning Prior to Sterilization:

• Prior to undergoing the sterilization process, soiled instruments/equipment are thoroughly cleaned using enzymatic detergent, rinsed, dried and inspected for the presence of dried blood or other debris. 

3a) (CE) Staff demonstrate /verbalize necessary steps/process to ensure sterility and/or high-level disinfection to ensure sterility/disinfection of equipment. 

• Personnel can demonstrate or verbally explain procedure(s) used for cleaning prior to sterilization, and to locate written directions on site. 

Cold Chemical Sterilization/High Level disinfection: 

• Product manufacturer’s directions are strictly followed for instrument pre-soaking treatment, solution preparation, solution exposure procedures, safety precautions (e.g., room temperature, area ventilation), and post-sterilization processes 

• Sterilization exposure times and solution expiration date/time are communicated to staff. 

• Written procedures for cold sterilization and /or high-level disinfection is available on site to staff 

Cold chemical sterilization/high level disinfection: 

3b) Confirmation from manufacturer item (s) is/are heat sensitive. 

• Per CDC, the use of liquid chemical germicides to sterilize instruments (“cold sterilization”) are limited. Sterility is not verified or assured with cold chemical sterilization. The first choice is always heat sterilization. The CDC refers to heat sterilization as “the method of choice when sterilizing instruments and devices. If an item is heat sensitive, it is preferable to use a heat-stable alternative or disposable item.” 

• The use of a liquid chemical sterilant should be restricted to reprocessing devices that are heat-sensitive and incompatible with other sterilization methods. All other items should be heat sterilized or disposable. 

• Product efficacy tests (i.e. test strips) shall be performed according to manufacturer’s guidelines 

Centers for Disease Control and Prevention: Disinfection and Sterilization 

https://www.cdc.gov/infectioncontrol/guidelines/disinfection/index.html

Centers for Disease Control and Prevention: Other Sterilization Methods: 

https://www.cdc.gov/infectioncontrol/guidelines/disinfection/sterilization/other-methods.html

Control Methods and Work Practices are in place to prevent or reduce exposure to the cold chemical sterilants. Cold chemical sterilants have toxic properties and are hazardous. 

• Cold chemical sterilants must be used strictly in accordance with the manufacturer’s directions. Always consult the manufacturer for safety precautions and MSDS information

• The appropriate PPE must be used to avoid inhalation or skin contact exposure to during the cold chemical sterilization/high level disinfection process 

Examples of chemicals include: 

o Glutaraldehyde (Cidex) 

o Peracetic acid 

o Hydrogen peroxide-based solutions 

Glutaraldehyde is a common cold chemical sterilants. Exposure to glutaraldehyde can cause the following health effects: throat and lung irritation, breathing difficulty, nose irritation, nosebleed, burning eyes and conjunctivitis, rash, hives, headaches, and nausea. 

Exposure to glutaraldehyde may be prevented or reduced by using the following control methods and work practices: 

o Use local exhaust ventilation 

o Keep glutaraldehyde baths under a fume hood where possible 

o Avoid skin contact (use appropriate PPE-gloves and aprons made of nitrile or butyl rubber wear goggles and face shields), 

o Use only enough sterilants to perform the required sterilization procedure, 

o Seal or cover all containers holding the sterilants 

o Attend training classes 

For more information on glutaraldehyde exposure and safety tips, refer to the Centers for Disease Control and Prevention: 

https://www.cdc.gov/niosh/docs/2001-115/default.html

3c) (CE) Appropriate PPE is available, exposure control plan, MSDS and clean up instructions in the event of a cold chemical sterilant spill. 

Cold Chemical Sterilants Spillage: The OSHA Hazard Communication Standard (29 CFR 1910.1200, 1915.99, 1917.28, 1918.90, 1926.59, and 1928.21) requires manufacturers and importers of hazardous chemicals to develop Material Safety Data Sheets (MSDS) for each chemical or mixture of chemicals. 

o Employers must have the data sheets for cold chemical sterilants readily available to employees who work with the products to which they could be exposed. 

o Staff should attend training classes in safety awareness about the use and exposure to cold chemical sterilants used on site. 

o Personnel are familiar with and can recognize signs and symptoms of exposure to cold chemical sterilants used on site.

o Staff must be aware of the procedures for clean up in the event of spillage. 

o Staff can demonstrate or verbally explain procedure(s) used on site for chemical spill cleanup 

o If personnel are unable to demonstrate or explain site-specific chemical spill cleanup procedures and cannot locate written chemical spill cleanup procedure instructions, site is considered deficient 

o Cleanup procedures may vary from site to site depending on the cold chemical sterilants used 

o The appropriate PPE for cold chemical sterilants clean up must be readily available. 

National Institute for Occupational Safety and Health (NIOSH) with the Centers for Disease Control and Prevention. Occupational Safety and Health Standards: 29 CFR 1910.1030 § d.3.i, ii; 4.ii.A, 4.iii.B, 1910.132, 1910.134. Environmental Health and Safety guidelines for disinfectants and sterilization methods. MSDS for cold chemical sterilants. The American National Standard (ANSI)/Advancing Safety in Medical Technology (AAMI) ST58:2013. 

https://www.cdc.gov/infectioncontrol/guidelines/disinfection/sterilization/index.html

4a) Staff demonstrate/verbalize necessary steps/process to ensure sterility. 

Autoclave/Steam Sterilization: 

• Autoclave manufacturer’s directions are strictly followed for instrument pre-cleaning, machine loading, operation safety precautions, minimum time-temperature criteria, and post sterilization processes 

• Written operating procedures for autoclave are available on site to staff 

• Documentation of sterilization loads include date, time and duration of run cycle, temperature, steam pressure, and operator of each run 

• If instruments/equipment are transported off-site for sterilization, equipment handling, and transport procedures are available on site to staff 

• Documentation of instruments and personnel transporting must be maintained 

4 b) Autoclave maintenance per manufacturer’s guidelines.

Autoclave Maintenance: Autoclave is maintained and serviced according to manufacturer’s guidelines. Documentation of maintenance should include: 

o Mechanical problems 

o Inspection dates 

o Results/outcome of routine servicing 

o Calibration 

o Repairs, etc. 

Note: If the manufacturer’s guidelines are not present on site, then the autoclave is serviced annually by a qualified technician. A dated sticker on the autoclave or a service receipt is acceptable documentation of appropriate maintenance. 

VI.D.4c) (CE) Spore testing of autoclave/steam sterilizer with documented results (at least monthly). 

Spore Testing: 

• Autoclave spore testing is performed at least monthly, unless otherwise stated in manufacturer’s guidelines. 

• Documentation of biological spore testing includes: 

o Date 

o Results 

o Types of spore test used 

o Person performing/documenting test results 

• Written procedures for performing routine spore testing and for handling positive spore test results are available on site to staff 

• For positive spore tests, the autoclave is removed from service immediately until inspection is completed and a negative retest occurs. Procedures include: Report/Repair/Retrieve/Retest/Re-sterilize: 

o Report problem 

o Repair autoclave 

o Retrieve all instruments sterilized since last negative spore test 

o Re-test autoclave 

o Re-sterilize retrieved instruments 

• Biologic spore test products vary and are designed for use based on specific autoclave type. Biologic control testing challenges the autoclave sterilization cycle with live, highly resistant, nonpathogenic spores. If spores are killed during processing, it is assumed that all other microorganisms are also killed and that the autoclave load is sterile.

Note: Documentation of monthly spore testing must be maintained onsite even for sterilization that is performed offsite. 

VI.D.4.d) (CE) Management of positive mechanical, chemical, and biological indicators of the sterilization process. 

Autoclave/Steam Sterilization Mechanical, Chemical, and Biological Indicators: 

• Sterilization failure can occur for reasons such as slight variation in the resistance of the spores, improper use of the sterilizer, and laboratory contamination during the culture. 

• Per CDC, the autoclave/steam sterilization procedure should be monitored routinely by using a combination of: 

o Mechanical: time: temperature, pressure in the sterilizer 

o Chemical: internal and external indicator on the package which suggest that the sterilizer was functioning properly 

o Biological: spore test – an indicator to evaluate the sterilizing conditions and indirectly the microbiologic status of the processed items 

• Staff should adhere to site-specific protocol and/or manufacturer/product label for management of positive indicator(s). 

https://www.cdc.gov/infectioncontrol/pdf/guidelines/disinfection-guidelines-H.pdf (Page 77) 

4.e) Sterilized packages are labeled with sterilization date and load identification information. 

Package and Storage of sterilized items: 

• Following the sterilization process, medical and surgical devices must be handled using aseptic technique in order to prevent contamination 

• Storage areas for sterilized packages are clean, dry and separated from non-sterile items by a functional barrier (e.g., shelf, cabinet door, and drawer) 

• Sterilized package labels include date of sterilization, load run identification information, and general contents (e.g. suture set) 

• Each item in a sterile package need not be listed on the label if a master list of package contents is available elsewhere on site 

4.f) Storage of sterilized packages. 

Storage of sterilized packages: 

• Storage areas for sterilized packages are clean, dry and separated from non-sterile items by a functional barrier (e.g., shelf, cabinet door, and drawer)

• Maintenance of sterility is event related, not time related

• Sterilized items are considered sterile until use, unless an event causes contamination 

• Sterilized items are not considered sterile if package is opened, wet/moist, discolored or damaged, and should be kept removed from sterile package storage area 

• Site has a process for routine evaluation of sterilized packages 