Clinical Services – Radiology Review

Site has current CA Radiologic Health Branch Inspection Report or Proof of Registration if there is radiological equipment on site. 

CDPH Radiologic Health Branch (RHB) Inspection Report: 

If site has current documentation of one of the following, give the full 9 points and survey items 2-9 will not need to be surveyed. Acceptable documentation is: 

o Inspection Report/Proof of Registration, or 

o Inspection Report/Proof of Registration and Short Form Sign-off sheet, or 

o Inspection Report/Proof of Registration and Notice of Violation form and approval letter for corrective action plan from the CA RHB 

The Radiologic Inspection Report or Proof of Registration (receipt of payment or cancelled check), issued by the RHB, must be present if there is radiology equipment on site. 

If any violations are found, one of two documents are issued to the site: 

o “Short Form Sign-off sheet” is issued for minimal problems that are easily corrected 

o “Notice of Violation” form, requiring a site corrective action plan, is issued if there are more violations that are serious. All “Notice of Violation” corrective action plans must be accompanied by an approval letter from the CA RHB. 

If documents are not available on site, or if reviewer is uncertain about the “status of documents on site, proceed to score all items 1-9. 

The following documents are posted on site: 

Current copy of Title 17 with a posted notice about availability of Title 17 and its location. 

“Radiation Safety Operating Procedures” posted in highly visible location. 

“Notice to Employees Poster” posted in highly visible location. 

“Caution, X-ray” sign posted on or next to door of each room that has X-ray equipment. 

Physician Supervisor/Operator certificate posted and within current expiration date. 

Technologist certificate posted and within current expiration date. 

The following radiological protective equipment is present on site: 

Operator protection devices: radiological equipment operator must use lead apron or lead shield. 

Gonadal shield (0.5 mm or greater lead equivalent): for patient procedures in which gonads are in direct beam. 

Radiological Equipment: 

Equipment inspection, based on a “priority” rating system, is established by legislation (CA H&S Code, Section 115115).

• Mammography equipment is inspected annually (Mammography Quality Standards Act, 21 CFR, Section 900), and must have federal FDA Certification on site and CA Mammography X-ray Equipment and Facility Accreditation Certification posted on the machine 

• High Priority equipment (e.g. fluoroscopy, portable X-ray) is inspected every three years 

• Medium Priority equipment is inspected every 4-5 years depending on the volume of patients, frequency of x-ray equipment uses, and likelihood of radiation exposure 

If reviewer is uncertain about the “status of equipment inspection, call the Radiological Health Branch. 

Radiology Personnel: 

• All certificates/licenses are posted and show expiration dates 

• If there are many technicians, a list of names, license numbers, and expiration dates may be substituted 

• The Certified Radiological Technologist (CRT) certificate permits the technologist to perform all radiology films except mammography and fluoroscopy, which require separate certificates 

• The “Limited Permit” restricts the technician to one of the ten-(10) x-ray categories specified on the limited certificate: Chest, Dental laboratory, Dermatology, Extremities, Gastrointestinal, Genitourinary, Leg-podiatric, Skull, Torso-skeletal, and X-ray bone densitometry 

Note: 

• Per RHB, dexascanners do not require lead aprons or gonadal shields, however, criteria 1-7 are still required 

• RHB uses the ALARA (As Low As Reasonably Achievable) principle, which is the foundation of all radiation safety programs. The ALARA principle means to minimize exposure to radiation doses by employing all reasonable methods 

• Dexascanners manufacturer guidelines do not require gonadal shielding or lead aprons due to very low radiation output, and potential for the shield to obscure the area being scanned, possibly rendering the scan non-diagnostic. With the focused beam, operators do not need aprons, the amount of exposure of “scattered” beams to an operator seated near the scanner is about the same level as that found in the natural environment 

A traditional x-ray machine used for bone density testing, is not a dexascanner, and may require shielding/apron. 

Note: The RHB of the Food, Drug, and Radiation Safety Division of the CA Department of Public Health (CDPH) enforces the Radiation Control Laws and Regulations designed to protect both the 

public and employees against radiation hazards. Enforcement is carried out through licensing, registration and periodic inspection of sources of radiation, such as radiation machines. 

For questions regarding radiologic safety (e.g. expired or no inspection letters on site), call CDPH Radiologic Health Branch at (916) 327-5106. 

For Radiation Emergency Assistance, call 1-800-852-7550. 

Ref: CCR, Title 17, Chapter 5, Subchapter 4 regulations at https://www.cdph.ca.gov/rhb