Facility Site Review Standards  

Purpose

The Facility Site Review Standards provide the instructions, rules, regulations, parameters, and indicators for conducting site reviews using the Facility Site Review tool. The site reviewer must use these Standards for measuring, evaluating, assessing, and making decisions. 

Scoring

Site reviews include on-site inspection and interviews with site personnel. Reviewers are expected to use reasonable evidence available during the review process to determine if practices and systems on site meet review criteria. Critical Elements have a weight of two (2) points each and non-Critical Elements have a weight of one (1) point on the site review tool. 

 Compliance levels include: 

  1. Exempted Pass: 90% or above without deficiencies in Critical Elements, Pharmaceutical or Infection Control 
  2. Conditional Pass: 80-89%, or 90% and above with deficiencies in either Critical Elements, Pharmaceutical or Infection Control 
  3. Fail: 79% and below 

 A corrective action plan (CAP) is required for a total score less than 90%, OR for a total score of 90% or above if there are deficiencies in Critical Elements, Pharmaceutical Services or Infection Control. Compliance rates are based on 169 total possible points, or on the total “adjusted” for Not Applicable (N/A) items. “N/A” applies to any scored item that does not apply to a specific site as determined by the reviewer. Reviewers are expected to determine how to ascertain information needed to complete the review. Review criteria to be reviewed only bya registered nurse (RN), nurse practitioner (NP), physician (MD), or physician assistant (PA) is labeled RN/NP/MD/PA.

Directions

Score full point(s) if review item is met. Score zero (0) points if item is not met. Do not score partial points for any item. Explain all “N/A” and “No” (0 point) items in the comment section. Provide assistance/consultation as needed for CAPs and establish follow-up/verification timeline. 

  1. Add the points given in each section. 
  2. Add points given for all six (6) sections to determine total points given for the site. 
  3. Subtract all “N/A” items from 169 total possible points to determine the “adjusted” total possible points. If there are no “N/A” items, calculation of site score will be based on 169 points. 
  4. Divide the total points given by 169 or by the “adjusted” total. Multiply by 100 to calculate percentage rate. 

Scoring Example

Step 1

Add the points given in each section. 

Step 2

Add points given for all six (6) sections. 

Example:

31 (Access/Safety) 

26 (Personnel) 

25 (Office Management) 

40 (Clinical Services) 

13 (Preventive Services) 

34 (Infection Control) 

____________

169 (POINTS GIVEN) 

Step 3

Subtract “N/A” points from 169 total points possible. 

169 (Total points possible) 

– 5 (N/A points) 

_____________

164 (“Adjusted” total points possible) 

Step 4

Divide total points given by the “adjusted” points, then multiply by 100 to calculate percentage rate. 

Points given 140
__________ ___________
“adjusted” total or 164 = 0.8537 x 100 = 85% 

Criteria for Access and Safety Standards

A. Site is accessible and useable by individuals with physical disabilities

Sites must have the following safety accommodations for physically disabled persons:

ADA Regulations
Site must meet city, county and state building structure and access ordinances for persons with physical disabilities. A site/facility includes the building structure, walkways, parking lots, and equipment 

All facilities designed, constructed; or altered by, on behalf of, or for the use of a public entity must be readily accessible and usable by individuals with disabilities, if the construction or alteration was begun after January 26, 1992 (28 CFR 35.151) 

Any alteration to a place of public accommodation or a commercial facility, after January 26, 1992, must be made to ensure that, to the maximum extent feasible, the altered portions of the facility are readily accessible to and useable by individuals with disabilities, including individuals who use wheelchairs (28 CFR 36.402)

A1) Clearly marked (blue) curb or sign designating disabled-parking space near accessible primary entrance. 

Parking:

• Parking spaces for persons with physical disabilities are located in close proximity to accessible building entrances 

• Each parking space reserved for persons with disabilities is identified by a permanently affixed reflectorized sign posted in a conspicuous place 

• If the provider has no control over availability of accessible parking within lot or nearby street spaces for persons with disabilities, the provider must have a plan in place for making program services available to persons with physical disabilities

A2) Pedestrian ramps have a level landing at the top and bottom of the ramp. 

Ramps: 

• A clear and level landing is at the top and bottom of all ramps and on each side of an exit door 

• Any path of travel is considered a ramp if its slope is greater than a 1-foot rise in 20 feet of horizontal run 

• Ramps must be a minimum of 36-inches wide. Some areas require wider ramps 

• All edges must be protected to keep anyone from slipping off 

• All ramps that are 5 feet long shall have a level top and bottom landings 

• Ramps must have handrails on both sides if length is longer than 6 feet 

A3) Exit and exam room doorway openings allow for clear passage of a person in a wheelchair.

Exit Doors: 

• All entrances and exterior and interior exit doors, regardless of the occupant load shall be made accessible to persons with disabilities 

• Exam room and exit doorways have a minimum opening of 32 inches with the door open at 90 degrees that will allow for passage of wheelchairs 

• Door hardware = operable with a single effort without requiring ability to grasp hardware 

• Effort to operate doors = 8.5 pounds at exterior doors and 5 pounds at interior doors 

• Door hardware height = 30” – 44” above floor 

• Exit doors include all doors required for access, circulation and use of the building and facilities, such as primary entrances and passageway doors 

• Furniture and other items do not obstruct exit doorways or interfere with door swing pathway 

A4) Accessible passenger elevator or reasonable alternative for multi-level floor accommodation. 

Elevators: 

• If there is no elevator, a freight elevator may be used to achieve program accessibility if it is upgraded for general passenger use and if passageways leading to and from the elevator are well-lit, neat and clean. 

A5) Clear floor space for wheelchair in waiting area and exam room. 

Clear Floor Space: 

• Clear space in waiting/exam areas is sufficient (at least 30-in. x 48-in.) to accommodate a single, stationary adult wheelchair and occupant 

• A minimum clear space of 60-inch diameter or square area is needed to turn a wheelchair 

Sanitary Facilities: 

A6) Wheelchair accessible restroom facilities. 

• A wheel-chair accessible restroom stall allows sufficient space for a wheelchair to enter and permits the door to close 

• Sufficient knee clearance space underneath the sink allows wheelchair users to safely use a lavatory sink for hand washing 

7) Wheelchair accessible handwashing facilities or reasonable alternative. 

• Restroom and hand washing facilities are accessible to able-bodied and physically disabled persons 

• If wheelchair-accessible restrooms are not available within the office site, reasonable alternative accommodations are provided such as a wheelchair-accessible restroom located within the building. 

• A reasonable alternative may include, but is not limited to, hand washing items provided as needed by site personnel e.g. urinal, bedpan or bedside commode in a private area

Note: 

• A public entity may not deny the benefits of its program, activities, and services to individuals with disabilities because its facilities are inaccessible (28 CFR 35.149-35.150) 

• Every feature need not be accessible, if a reasonable portion of the facilities and accommodations provided is accessible (Title 24, Section 2-419, California Administrative Code, the State Building Code) 

• Reasonable Portion and/or Reasonable Alternatives are acceptable to achieve program accessibility 

• Reasonable Portion applies to multi-storied structures and provides exceptions to the regulations requiring accessibility to all portions of a facility/site 

• Reasonable Alternatives are methods other than site structural changes to achieve program accessibility, such as acquisition or redesign of equipment, assignment of assistants/aides to beneficiaries, provision of services at alternate accessible sites, and/or other site-specific alternatives to provide services (ADA, Title II, 5.2000) 

• Points shall not be deducted if Reasonable Portion or Reasonable Alternative is made available on site. 

Specific measurements are provided strictly for “reference only” for the reviewer. Site reviewers are NOT expected to measure parking areas, pedestrian path of travel walkways and/or building structures on site.

B. Site environment is maintained in a clean and sanitary condition.

B1) All patient areas including floor/carpet, walls, and furniture are neat, clean, and well maintained. 

• The physical appearance of floors/carpets, walls, furniture, patient areas and restrooms are clean and well maintained 

B2) Restrooms are clean and contain appropriate sanitary supplies. 

• Appropriate sanitary supplies, such as toilet tissue, hand washing soap, cloth/paper towels or antiseptic towelettes are made available for restroom use 

• Environmental safety includes the “housekeeping” or hygienic condition of the site 

• Clean means unsoiled, neat, tidy, and uncluttered 

• “Well maintained” means being in good repair or condition 

C. Site environment is safe for all patients

Ordinances: 

• Sites must meet city, county, and state fire safety and prevention ordinances 

• Reviewers should be aware of applicable city and county ordinances in the areas in which they conduct reviews

There is evidence staff has received safety training and/or has safety information available on the following: 

C1) Fire safety and prevention. 

C2) Emergency non-medical procedures (e.g. site evacuation, workplace violence). 

Emergency Action Plans: 

• Non-medical emergencies include incidents of fire, natural disaster (e.g. earthquakes), workplace violence, etc. 

• Specific information for handling fire emergencies and evacuation procedures is available on site to staff. Personnel know where to locate information on site, and how to use information. (29 CFR 1910.38) 

C3) Lighting is adequate in all areas to ensure safety. 

Illumination: Lighting is adequate in-patient flow working and walking areas such as corridors, walkways, waiting and exam rooms, and restrooms to allow for a safe path of travel 

C4) (CE) Exit doors and aisles are unobstructed and egress (escape) accessible. 

Access Aisle: 

• Accessible pedestrian paths of travel (ramps, corridors, walkways, lobbies, elevators, etc.) between elements (seats, tables, displays, equipment, parking spaces, etc.) provide a clear circulation path 

• Means of egress (escape routes) are maintained free of obstructions or impediments to full instant use of the path of travel in case of fire or other type of emergency 

• Building escape routes provide an accessible, unobstructed path of travel for pedestrians and/or wheelchair users at all times when the site is occupied 

• Cords (including taped cords) or other items are not placed on or across walkway areas 

C5) Exit doors are clearly marked with “Exit” signs. 

Exits: Exit doorways are unobstructed and clearly marked by a readily visible “Exit” sign (29 CFR 1910.37)

C6) Clearly diagramed “Evacuation Routes” for emergencies are posted in a visible location at all elevators, stairs and exits. 

Evacuation Routes: 

• Clearly marked, easy-to-follow escape routes are posted in visible areas, such as hallways, exam rooms and patient waiting areas, as well as posted on all elevators, stairs and exits 

• The minimum clear passage needed for a single wheelchair is 36 inches along an accessible route but may be reduced to a minimum of 32 inches at a doorway 

C7) Electrical cords and outlets are in good working condition. 

Electrical Safety: 

• Electrical cords are in good working condition with no exposed wires, frayed or cracked areas Cords are not affixed to structures, placed in or across walkways, extended through walls, floors, and ceiling or under doors or floor coverings 

• Extension cords are not used as a substitute for permanent wiring 

• All electrical outlets have an intact wall faceplate 

• Sufficient clearance is maintained around lights and heating units to prevent combustible ignition 

C8) Fire Fighting Equipment in accessible location. 

Firefighting equipment: 

There is firefighting equipment that must be in accessible locations on site. At least one of the following types of fire safety equipment is on site: 

o Fire Extinguisher: The employer shall provide portable fire extinguishers and shall mount, locate and identify them so that they are readily accessible. Fire extinguishers are maintained in a fully charged and operable condition and kept in their designated places at all times except during use(29 CFR 1910.157) 

o Smoke Detector with intact batteries 

o Automatic Sprinkler System With a 10-inch clearance between sprinkler heads and stored materials 

https://www.osha.gov/laws-regs/regulations/standardnumber/1910/1910.157

C9) An employee alarm system. 

Employee Alarm System: 

• Employers must install and maintain an operable employee alarm system that has a distinctive signal to warn employees of fire or other emergencies, unless employees can promptly see or smell a fire or other hazard in time to provide adequate warning to them. (29 CFR 1910. 37) 

OSHA: For those employers with 10 or fewer employees in a workplace, direct voice communication is an acceptable procedure for sounding the alarm provided all employees can hear the alarm. Such workplaces do not need a back-up system. 

https://www.osha.gov/laws-regs/regulations/standardnumber/1910/1910.37

Note: Specific measurements are provided strictly for “reference only” for the reviewer. Site reviewers are NOT expected to measure parking areas, pedestrian path of travel walkways and/or building structures on site.

D. Emergency health care services are available and accessible 24 hours a day, 7 days a week.

D1) Personnel are trained in procedures/action plan to be carried out in case of medical emergency on site.

Site Specific Emergency Procedures: 

• Staff can describe site-specific actions or procedures for handling medical emergencies until the individual is stable or under care of local emergency medical services (EMS) 

• There is a written procedure for providing immediate emergent medical care on site until the local EMS is on the scene. Although site proximity to emergency care facilities may be considered when evaluating medical emergency procedures, the key factor is the ability to provide immediate care to patients on site until the patient is stable or EMS has taken over care/treatment 

• When the MD or NPMP is not on site, staff/MA may call 911, and CPR-certified staff may initiate CPR if needed 

• Non-CPR-certified staff may only call 911 and stay with the patient until help arrives 

D2) Emergency equipment is stored together in easily accessible location and is ready to be used. 

Emergency Medical Equipment: 

During business hours providers are prepared to provide emergency services for management of emergency medical conditions that occur on site until the emergent situation is stabilized and/or treatment is initiated by the local 911 Emergency Medical Service (EMS) system. Minimum emergency equipment is available on site to: 

o Establish and maintain a patent/open airway

o Manage emergency medical conditions 

Emergency equipment and medication, appropriate to patient population served, are available in an accessible location and is ready for use. 

• An accessible location is one that is reachable by personnel standing on the floor, or other permanent working area, without locating/retrieving step stool, ladder or other assistive devices 

• For emergency “Crash” cart/kit, contents are appropriately sealed and are within the expiration dates posted on label/seal 

• Site personnel are appropriately trained and can demonstrate knowledge and correct use of all medical equipment they are expected to operate within their scope of work 

• Documented evidence that emergency medication and equipment is checked at least monthly may include a log, checklist or other appropriate method(s) 

https://www.aafp.org/afp/2007/0601/p1679.html

D3) Emergency phone number contacts are posted, updated annually and as changes occur. 

Emergency Phone Number list: 

Posted in an accessible and prominent location(s) and includes: 

o Local emergency response services (e.g., fire, police/sheriff, ambulance) 

o Emergency contacts (e.g., responsible managers, supervisors) 

o Appropriate State, County, City and local agencies (e.g., local poison control number)

Emergency medical equipment appropriate to practice/patient population is available on site: 

D4) Airway management: oxygen delivery system, nasal cannula or mask, bulb syringe and Ambu bag. 

Airway Management: 

Without the ability to adequately maintain the patient’s airway, all other interventions are futile. Minimum airway control equipment with various sizes of airway devices appropriate to patient population within the practice includes: 

o Wall oxygen delivery system or portable oxygen tank 

o Nasal cannula or mask 

o Bulb syringe 

o Ambu bag as appropriate to patient population served. Mask should be replaced when they no longer make a solid seal 

o Portable oxygen tanks are maintained at least ¾ full. There is a method/system in place for oxygen tank replacement. If oxygen tanks are less than ¾ full at time of site visit, site has a back-up method for supplying oxygen if needed and a scheduled plan for tank replacement. 

o Oxygen tubing does not need be connected to oxygen tank, but must be kept in close proximity to tank 

Oropharyngeal airways are no longer required. 

D5) (CE) Emergency medicine such as asthma, chest pain, hypoglycemia and anaphylactic reaction management: 

Emergency Medication\Anaphylactic Reaction Management: 

Severe allergic reaction can cause urticaria (hives), hypotension, bronchospasm, wheezing, and pulmonary edema. Per American Academy of Family Practice (AAFP) asthma exacerbation, chest pain, hypoglycemia management, minimum equipment based on the patient population served includes: 

o Epinephrine 1 mg/mL (injectable) 

o Diphenhydramine 25 mg (oral) or 50 mg/ml (injectable) 

o Naloxone 

o Chewable aspirin 81 mg 

o Nitroglycerin spray/tablet* 

o Bronchodilator medication (solution for nebulizer or metered dose inhaler) 

o Glucose (any type of glucose) 

o Appropriate sizes of ESIP needles/syringes** and alcohol wipes 

• The typical adult strength to address cardiac emergencies is 325 mg (four doses of 81 mg chewable aspirin or one dose of 325 non-enteric coated aspirin) 

• If the site is seeing adults, the reviewer shall assess whether the appropriate number of chewable aspirin tablets of 81 mg is available (at least four tablets) 

*Pediatric offices only serving patients under 18 years old are not required to keep Nitroglycerin in their emergency kit. According to the FDA, “The safety and effectiveness of nitroglycerin in pediatric patients (under 18 years old) have not been established.” 

**If the emergency kit or “crash cart” has only non-safety needles/syringes, score that deficiency in Section VI., Infection Control, criteria B.2. See Infection Control Standards. 

https://www.fda.gov/drugs/postmarket-drug-safety-information-patients-and-providers/narcan-naloxone-nasal-spray-approved-reverse-opioid-overdose

https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5753997/

https://www.heart.org/en/health-topics/heart-attack/treatment-of-a-heart-attack/aspirin-and-heart-disease

https://www.accessdata.fda.gov/drugsatfda_docs/label/2014/021134s007lbl.pdf

D6) Medication dosage chart for all medications included with emergency equipment (or other method for determining dosage) is kept with emergency medications. 

• There is a current medication administration reference (e.g. medication dosage chart) available for readily identifying the correct medication dosages (e.g. adult, pediatric, infant, etc.) 

• Package inserts are not acceptable as dosage charts 

• All emergency medications in the emergency kit/ crash cart must have dosage charts 

• A receipt or documentation showing medication is ordered is acceptable for any medication shortage 

• Score should be either a Yes or No only 

There is a process in place on site to: 

D7) Document checking of emergency equipment/supplies for expiration and operating status at least monthly. 

D8) Replace/re-stock emergency medication, equipment and supplies immediately after use. 

Note: An “emergency medical condition” is a medical condition that manifests itself by acute symptoms of sufficient severity (including severe pain) such that the absence of immediate medical attention could reasonably be expected to result in:

1) placing the health of the individual (or unborn child of a pregnant woman) in serious jeopardy 

2) serious impairment to bodily functions 

3) serious dysfunction of any bodily organ or part 

“Emergency services” means those services required for alleviation of severe pain, or immediate diagnosis and treatment of unforeseen medical conditions, which, if not immediately diagnosed and treated, would lead to disability or death.

E. Medical and lab equipment used for patient care is properly maintained. RN/NP/MD/PA

E1) Medical equipment is clean

Medical and Laboratory Equipment: 

All equipment used to measure or assess patient health status/condition is clean. 

E2) Written documentation demonstrates the appropriate maintenance of all medical equipment according to equipment manufacturer’s guidelines. 

Documentation: 

• There is documented evidence that standard operating procedures have been followed for routine inspection/maintenance, calibration, repair of failure or malfunction, and testing and cleaning of all specialized equipment 

• Appropriate written records include calibration or other written logs, work orders, service receipts, dated inspection sticker, etc. 

• All equipment used to measure or assess patient health status/condition is functioning properly. All specialized equipment (e.g., ultrasonography equipment, electrocardiogram (EKG) machine, defibrillator, audiometer, hemoglobin meter, glucometer, scales, etc.) are adequately maintained according to the specified manufacturer’s guidelines for the equipment, or is serviced annually by a qualified technician 

• Blood pressure cuffs, monitors, and other related equipment need not be calibrated unless required by the manufacturer. Manufacturer guidelines must be available on site, indicating that it is not necessary to calibrate the equipment

Note: The term monitor includes, but not limited to, glucometers, EKG, BP monitors, hemocues, and audiometers.